Changes for page PID
Last modified by TerraIndex TerraIndex on 2022/04/13 10:10
From version 6.2
edited by TerraIndex TerraIndex
on 2021/05/28 11:21
on 2021/05/28 11:21
Change comment:
Added tag [PetroFLAG analyzer, PID]
To version 5.1
edited by TerraIndex TerraIndex
on 2021/05/28 11:20
on 2021/05/28 11:20
Change comment:
Uploaded new attachment "741px-Potentials.png", version {1}
Summary
-
Page properties (2 modified, 0 added, 0 removed)
Details
- Page properties
-
- Tags
-
... ... @@ -1,1 +1,0 @@ 1 -PetroFLAG analyzer|PID - Content
-
... ... @@ -18,7 +18,6 @@ 18 18 19 19 To activate the gas an electric charge is needed to extract an electron from the gas. This charge varies with the type of gas present and is expressed in Electron Volts. The minimal amount of electron volt needed is called the ionization potential. With the PID-meter one has three choices the 9.8 eV, the 10,6 eV and the 11.7 eV lamp. The higher the amount of charge the more types of gasses can be detected. On the right there is a graphical display of the type of gasses and their activation charge. 20 20 21 -[[image:741px-Potentials.png||height="245" width="407"]] 22 22 23 23 Apart from this ionization potential most of the times also a correction factor is known. This factor can be used when one suspects that only one type of gas is present in the field. By applying the correction factor on the PID-meter the concentration detection becomes more accurate. If the correction fact is not used the PID-meter does a wide spectrum measurement calibrated with isobutylene. 24 24