Changes for page PetroFLAG analyzer

Last modified by TerraIndex TerraIndex on 2021/09/30 15:20

From version 6.1
edited by TerraIndex TerraIndex
on 2021/07/16 13:27
Change comment: There is no comment for this version
To version 3.1
edited by TerraIndex TerraIndex
on 2021/05/28 11:45
Change comment: There is no comment for this version

Summary

Details

Page properties
Tags
... ... @@ -1,1 +1,0 @@
1 -PID|PetroFLAG analyzer
Content
... ... @@ -20,6 +20,7 @@
20 20  
21 21  === Response factors and method detection limits for common hydrocarbons ===
22 22  
23 +(% class="table-hover" %)
23 23  |=Hydrocarbon type|=Method detection limit (ppm)|=Response setting
24 24  |Transformer oil|15|10
25 25  |Grease|15|9
... ... @@ -137,7 +137,7 @@
137 137  
138 138  If you wish to exit the read mode, push the <NEXT> key and the operation is returned to the MAIN MENU. The <NEXT> key is also used to skip a step where a menu selection is required. To change a flashing menu option, push the <SCROLL> key while the option is flashing. To store the currently flashing menu choice, push the <SELECT> key. This stores the current choice and moves the flashing cursor to the next program mode.[[[2]>>url:https://www.equipcoservices.com/pdf/manuals/dexsil_petroflag_system.pdf]]
139 139  
140 -== Choosing a calibration curve ==
141 +== Selecting a calibration curve ==
141 141  
142 142  Either of the two calibration curves, identified as "1C" and "2C", can be selected from the MAIN MENU. From either calibration curve any response factor can be selected. To change the response factor or to recalibrate the unit, use the <NEXT> key to enter the MAIN MENU screen. Immediately upon entering this menu three decimal points and the response factor are displayed. Next, the first two characters on the screen indicates the calibration curve that is currently selected (“1C” or “2C”) is displayed. They will blink, indicating that a new curve may be selected. Use the <SCROLL> key to scroll to the next calibration curve. Push the <SELECT/OFF> key to select the curve. The response factor will then blink. Use the <“SCROLL> key to scroll to the desired response factor for the target analyte and press the <SELECT/OFF> key.[[[2]>>url:https://www.equipcoservices.com/pdf/manuals/dexsil_petroflag_system.pdf]]
143 143  
... ... @@ -155,7 +155,7 @@
155 155  
156 156  = Errors =
157 157  
158 -== Error conditions ==
159 +=== Error conditions ===
159 159  
160 160  (% class="table-hover" %)
161 161  |=Message|=Cause|=Solution