... ... @@ -1,0 +1,1 @@
1 +PID|PetroFLAG analyzer
Content
... ... @@ -20,7 +20,6 @@
20 20  
21 21  === Response factors and method detection limits for common hydrocarbons ===
22 22  
23 -(% class="table-hover" %)
24 24  |=Hydrocarbon type|=Method detection limit (ppm)|=Response setting
25 25  |Transformer oil|15|10
26 26  |Grease|15|9
... ... @@ -138,7 +138,7 @@
138 138  
139 139  If you wish to exit the read mode, push the <NEXT> key and the operation is returned to the MAIN MENU. The <NEXT> key is also used to skip a step where a menu selection is required. To change a flashing menu option, push the <SCROLL> key while the option is flashing. To store the currently flashing menu choice, push the <SELECT> key. This stores the current choice and moves the flashing cursor to the next program mode.[[[2]>>url:https://www.equipcoservices.com/pdf/manuals/dexsil_petroflag_system.pdf]]
140 140  
141 -== Selecting a calibration curve ==
140 +== Choosing a calibration curve ==
142 142  
143 143  Either of the two calibration curves, identified as "1C" and "2C", can be selected from the MAIN MENU. From either calibration curve any response factor can be selected. To change the response factor or to recalibrate the unit, use the <NEXT> key to enter the MAIN MENU screen. Immediately upon entering this menu three decimal points and the response factor are displayed. Next, the first two characters on the screen indicates the calibration curve that is currently selected (“1C” or “2C”) is displayed. They will blink, indicating that a new curve may be selected. Use the <SCROLL> key to scroll to the next calibration curve. Push the <SELECT/OFF> key to select the curve. The response factor will then blink. Use the <“SCROLL> key to scroll to the desired response factor for the target analyte and press the <SELECT/OFF> key.[[[2]>>url:https://www.equipcoservices.com/pdf/manuals/dexsil_petroflag_system.pdf]]
144 144  
... ... @@ -156,7 +156,7 @@
156 156  
157 157  = Errors =
158 158  
159 -=== Error conditions ===
158 +== Error conditions ==
160 160  
161 161  (% class="table-hover" %)
162 162  |=Message|=Cause|=Solution